High severity · published Mar 13, 2025
CVE-2025-2229
Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
7.7
CVSS
0.1%
EPSS
No
On KEV
Description
A token is created using the username, current date/time, and a fixed
AES-128 encryption key, which is the same across all installations.
CVSS vector
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Attack vector
Local
Complexity
Low
Privileges
None
User interaction
None
Confidentiality
High
Integrity
High
Availability
None
Scope
Unchanged
Exploitation
Is it being used?
Exploit prediction (EPSS)0.1% · higher than 2% of CVEs
Exploited in the wildNo evidence yet
CISA KEVNot listed
Ransomware campaign useNot known
Affected products
Which versions
IC
Intellispace Cardiovascular (ISCV) · Philips
through 4.1 · source: vendor
Affected product
IC
Intellispace Cardiovascular (ISCV)
Philips
References
NVD
https://nvd.nist.gov/vuln/detail/CVE-2025-2229
CVE.org
https://www.cve.org/CVERecord?id=CVE-2025-2229
www.cisa.gov
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01
www.philips.com
https://www.philips.com/a-w/security/security-advisories.html
More from Intellispace Cardiovascular (ISCV)
7.7
CVE-2025-2230
Philips Intellispace Cardiovascular (ISCV) Improper Authentication